Product Introduction
The TESTech Photovoltaic Module Flammability Tester is designed and developed in accordance with ISO 11925-2:2010. It is used to determine the ignitability of building products by applying a small flame directly to vertically mounted specimens without additional radiant heat.
To accommodate photovoltaic module specimens, the apparatus is equipped with a customized large combustion chamber, an adjustable specimen mounting fixture, and an exhaust system.
The maximum specimen size is 1,200 mm wide × 2,200 mm high. Specimens can be conveniently loaded through the front of the chamber.
Applicable Standards
| Standard | Description |
|---|---|
| ISO 11925-2:2010 | Reaction-to-fire tests: ignitability of products subjected to direct flame impingement |
| IEC 61730-2:2016 | Photovoltaic module safety qualification: testing requirements |
Structural Features
1. The chamber has a painted exterior finish, providing an attractive and durable appearance.
2. Chamber dimensions:
Internal dimensions: 1,500 mm (L) × 800 mm (W) × 2,700 mm (H)
Door opening: 1,390 mm × 2,090 mm
3. Specimens are loaded through the front door. The mounting frame can be tilted forward for specimen installation and then easily returned to the upright test position and secured inside the chamber.
4. The fixture can accommodate specimens up to: 1,200 mm * 2,200 mm
The specimen fixture is adjustable for different specimen dimensions.
5. Ignition positions:
Surface ignition position: 40 mm above the bottom edge
Bottom-edge ignition position: 1.5 mm from the front surface at the specimen's lower edge
6. The default flame-application time is 15 seconds, although any required exposure time can be programmed.
7. A reinforced stainless-steel specimen fixture is installed inside the chamber. The fixture can be tilted outside the front door for specimen installation and then rotated back into the combustion chamber.
A bottom-edge ignition opening is provided at the base of the fixture. The opening has a width of 200 mm.
8. The burner can be moved forward, backward, upward, and downward to adjust its position relative to the specimen.
Because photovoltaic specimens can be relatively heavy, the apparatus uses a movable burner rather than requiring the specimen to be repositioned. This makes test preparation easier and reduces manual handling.
9. The burner can move freely in the following directions:
Forward and backward
Left and right
Up and down
This enables both bottom-edge and surface ignition tests.
10. Because of the large size of the apparatus, a handheld timing device is included. This allows the operator to observe and time the test while away from the main control panel.
11. An explosion-protected internal lamp illuminates the combustion chamber.
12. The imported burner has:
An internal nozzle diameter of 0.17 mm
Four air-adjustment holes, each with a diameter of 4 mm
13. An imported Japanese gas-flow control valve provides precise adjustment of the flame height to: 20+-1 mm
14. A pressure gauge and pressure-regulating valve provide accurate control of the gas pressure.
15. The flame height is adjustable, and automatic ignition is provided.
16. The flame-application time can be freely programmed. When the control program starts, the default ignition time is set according to IEC 61730-2.
17. Gas-pressure range: 10–50kPa
18. Available burner angles:
0°, vertical
45°
19. A tempered-glass door allows the operator to observe the test safely.
20. All internal fixtures are made of stainless steel for corrosion resistance and a clean appearance.
21. The apparatus includes the specimen holder and test program introduced in ISO 11925-2:2010.
22. The system provides a high degree of automation, accurate test data, and straightforward operation.
Control System
The apparatus uses a programmable logic controller (PLC) and touchscreen control system.
Its principal functions include:
Automatic timing
Automatic ignition
Ignition-failure detection and protection
Storage of test data for up to six specimens
LCD touchscreen display
Test-report retrieval
Data export through USB
Real-time monitoring of test data
Retrieval of results for six specimens
Alarm-history display
Audible alarms for abnormal test conditions
Adjustable exhaust-fan speed
At a position 50 mm from the specimen surface, the airflow velocity is controlled as follows:
| Airflow Direction | Maximum Velocity |
|---|---|
| Vertical | <0.2 m/s |
| Horizontal | <0.1 m/s |
Product Specifications
| Item | Specification |
|---|---|
| Power supply | 220 V, 50/60 Hz, 10 A |
| External dimensions | 2,100 mm (L) × 900 mm (W) × 3,260 mm (H) |
| Required installation space | 3,100 mm (L) × 1,900 mm (W) × 3,500 mm (H) |
| Maximum specimen size | 1,200 mm (W) × 2,200 mm (H) |
| Equipment weight | Approximately 500 kg |
| Fuel gas | Propane, supplied by the customer |
| Gas-pressure range | 10–50 kPa |
| Standard flame height | 20 ± 1 mm |
| Burner angle | 0° or 45° |
| Default flame-application time | 15 seconds, programmable |