TESTech Instrument (Suzhou) Technologies is one of the leading manufacturers and suppliers of gasmet ftir system for toxicity test in fire testing application in China. Welcome to wholesale premium test equipment at competitive price from our factory. If you have any enquiry about quotation, please feel free to email us.
Product Introduction
The Fourier transform infrared spectrometer is an ideal tool for toxicity testing in fire testing applications. It can be used together with a cone calorimeter or a smoke density chamber to perform online detection of toxic gas components released during combustion.
Typical smoke toxicity components measured in CEM applications include H2O, CO2, CO, N2O, NO, NO2, SO2, HCl, HF, NH3, CH4, C2H6, C3H8, and C2H4. The Gasmet FTIR spectrometer can be easily expanded with new test components and ranges. The measured components and calibration ranges can be changed according to actual application needs, and the technical configuration data are contained in one sampling system.
The FTIR spectrometer is used for online measurement and is ideal for determining trace gas concentrations in combustion tests. All parts of the FTIR analytical system are heated to 180 C. It can analyze both diluted gas and sample gas without pre-drying.
The Gasmet FTIR analytical system consists of the FTIR CX4000 gas analyzer meeting relevant ISO 19702 requirements, the Gasmet industrial PC system, and the Gasmet oxygen sampling system. All relevant components are integrated into a standard 19-inch cabinet.
The system includes power connectors and heated components for the sample line and sampling probe. It runs automatically and is controlled by Calcmet software. Measurement data can be transmitted from the industrial PC to the control room through digital output (ModBus) or 4-20 mA analog output.
The system provides different alarm functions, including function alarms, service alarms, and system alarms. When any alarm signal is activated, instrument air is automatically introduced to prevent condensation.
Functional Technical Parameters
- Fixed fully automatic continuous measurement system.
- Mid-infrared full-spectrum analysis, spectral range 500 to 4500 cm-1.
- Resolution 4 cm-1, interferometer scanning speed 10 scans per second.
- Detector: Peltier-cooled MCT; infrared source: SiC150K; beamsplitter: ZnSe moisture-resistant material.
- Measurement accuracy: ±2 percent of full scale.
- Operating temperature 180 C, multi-reflection optical path 5.0 m.
- Dust filtration above class 2, 2 μm.
- Moisture-resistant optical window materials suitable for humid working environments.
- Continuous long-term data testing, continuous recording and storage of spectra and data, and historical data playback.
- Gas cell materials use gold coating and rhodium coating to resist corrosive gases such as HCl, HCN, and NH3.
- Can measure mixed gases with H2O up to 40 percent and permanently store the spectral library. After factory gas component calibration, recalibration is not required.
- Response time T90 <120 s.
- Full heated-path constant temperature control.
- Automatic temperature and pressure compensation with user-friendly operation.
- Low maintenance cost, no sample loss or change, fully automatic testing system, complete safety functions, and fully modular system for maximum flexibility.
- Before the sampling unit, no dilution or dehydration of the sample gas is required; all parts are heated to 180 C.
- Industrial PC operation.
- If the sampler encounters an emergency alarm, instrument air automatically purges the system.
- Span and zero check valves operate automatically.
- Uses the CLS calculation method.
- Can be connected to combustion test equipment for online continuous measurement of toxic gas concentration in smoke.
- Fully meets ISO 19702 requirements.
General Parameters
|
Parameter |
Value |
|
Measurement Principle |
FTIR, Fourier transform infrared |
|
Performance |
Can simultaneously measure more than 50 gas components |
|
Operating Temperature |
20 ±20 C, non-condensing, dust-free environment |
|
Storage Temperature |
-20 to +60 C |
|
Response Time T90 |
<120 s, depending on gas flow and detection time |
|
Gas Analysis Unit Temperature |
50 to 180 C |
|
Sample Gas |
Non-condensing gas without dust particles |
|
Gas Flow Rate |
~4 l/min |
|
Sample Gas Pressure |
Atmospheric pressure |
Measuring Parameters
|
Parameter |
Value |
|
Zero Calibration |
Every 24 h with pure N2, 5.0 grade or higher recommended |
|
Zero Drift |
<2 percent, less than 2 percent of each zero calibration interval range |
|
Sensitivity Drift |
None |
|
Linearity Deviation |
<2 percent of measuring range |
|
Temperature Drift |
Drift <2 percent of range for each 10 K temperature change |
|
Pressure Influence |
A 1 percent change in sample gas pressure causes a 1 percent change in measured value; ambient pressure changes can be monitored and compensated |
Alarm Outputs
|
Parameter |
Value |
|
Function Alarm |
Gasmet FTIR analyzer and Calcmet application software |
|
Service Alarm |
Gasmet FTIR analyzer and Calcmet application software |
|
System Alarm |
Heating module temperature low/high, low flow alarm, low zero-gas pressure alarm |
Measuring Data Outputs
|
Parameter |
Value |
|
Output Control |
Analog or digital output configured and controlled by Gasmet industrial PC |
|
Digital Output |
ModBus, ASCII, COMLI, DDE, and other required connections |
|
Analog Output |
4-20 mA independent output |
|
Channels |
8 or 16, 1 or 32 bit PCI board |
Heated Lines
|
Parameter |
Value |
|
Tube Size |
4/6 mm |
|
Core Material |
PTFE |
|
Operating Pressure |
Maximum 400 kPa |
|
Temperature |
Maximum 200 C |
|
Accessories |
6 mm fitting sleeve |
|
Power Supply |
230 VAC or 115 VAC |
|
Power Density |
120 W/m |
|
Maximum Heated Line Length |
30 m at 230 VAC, 15 m at 115 VAC |
|
Interconnection |
All analysis units connected by 1 m heated lines |
Sampling Probe
|
Parameter |
Value |
|
Probe Type |
PSP4000H |
|
Power |
800 W |
|
Operating Temperature |
180 C |
|
Filter Unit |
Ceramic, 2 μm |
|
Dust Concentration |
<2 g/m3 |
|
Sampling Tube Material |
316 stainless steel |
|
Probe Length |
1 m |
|
Sample Gas Temperature |
Up to 600 C |
|
Sample Gas Pressure |
0.4 to 6 bar |
|
Mounting Flange |
DN65PN6 |
Electrical Connections
|
Parameter |
Value |
|
Mains Supply |
3 x 16 A, 3 x L+N+PE |
|
Power Consumption |
~7.5 kW including complete Gasmet FTIR CEMS groups and all components: FTIR gas analyzer, industrial PC, sampling unit, FTIR O2 analyzer, sample probe, and 21 m heated line |
Factory-Calibrated Components and Ranges
|
Component |
Calibration Range |
|
H2O |
0 to 40 vol% |
|
CO |
0 to 10000 ppm |
|
NO |
0 to 2000 ppm |
|
SO2 |
0 to 500 ppm |
|
NH3 |
0 to 1000 ppm |
|
HCl |
0 to 500 ppm |
|
HF |
0 to 400 ppm |
|
HBr |
0 to 100 ppm |
|
CO2 |
0 to 20 vol% |
|
NO2 |
0 to 1000 ppm |
|
N2O |
0 to 1000 ppm |
|
CH4 |
0 to 500 ppm |
|
C2H4 |
0 to 500 ppm |
|
C2H6 |
0 to 500 ppm |
|
C3H8 |
0 to 500 ppm |
|
C6H14 |
0 to 500 ppm |
|
HCN |
0 to 200 ppm |
|
CH2O |
0 to 200 ppm |
|
C3H6O |
0 to 200 ppm |
|
C3H4O |
0 to 200 ppm |